Keysight4142 B
No photo available yet

What is it?

The Keysight / Agilent 4142B is a modular semiconductor parameter tester mainframe with eight slots for plug-in measurement units.

What do the numbers mean?

Power & electrical12

Integrated measurement functions
Current/voltage sources, nanovoltmeters, and picoammeters in one instrument[1]
Accurate?
Source Monitor Unit (SMU) sensitivity
Forces voltage while measuring currents down to 20 fA; forces current while monitoring voltages down to 40 µV[1]
Accurate?
HP 41420A High Power SMU
2 slots; 40 µV to 200 V, 20 fA to 1 A (±200 V and ±1 A)[1]
Accurate?
HP 41421B Medium Power SMU
1 slot; 40 µV to 100 V, 20 fA to 100 mA (±100 V and ±100 mA)[1]
Accurate?
HP 41422A High Current Unit (HCU)
2 slots; pulse mode only; 100 µs pulsed measurements up to ±10 V and 10 A[1]
Accurate?
HP 41423A High Voltage Unit (HVU)
2 slots; DC or pulsed measurements to ±1000 V; low-leakage design measures currents down to 2 pA[1]
Accurate?
HP 41425A Analog Feedback Unit (AFU)
1 slot; used with two SMUs for analog voltage/current searches; finds target values in 12 ms[1]
Accurate?
Maximum configuration coverage
2000 V max using two HVUs; 20 A (pulsed) using two HCUs[1]
Accurate?
Maximum common-to-ground voltage
±42 V[1]
Accurate?
Ground Unit (GNDU)
High-current SMU set to 0 V, used as measurement ground; Kelvin connection; maximum offset voltage ±500 µV; current range ±1.6 A[1]
Accurate?
Power requirements
100/120/220 V ±10%; 240 V -10%/+5%; maximum 750 VA (48-66 Hz)[1]
Accurate?
Typical measurement time (SMU)
HP 41420A/HP 41421B SMU (20 V/100 mA range, spot measurement): force I or V in 3.5 ms; measure I or V in 4.0 ms[1]
Accurate?

Control & software1

Interfaces
HP-IB; external trigger input (TTL, negative logic, minimum pulse width 100 µs); external trigger output (TTL, negative logic)[1]
Accurate?

Dimensions & facilities2

Dimensions
426 mm W × 235 mm H × 676 mm D[1]
Accurate?
Weight
Mainframe approx. 23 kg; plug-in units 2-3 kg each[1]
Accurate?

Configuration & options9

Architecture
Mainframe with 8 slots for plug-in measurement units[1]
Accurate?
Maximum output power
Sum of all measurement units' instantaneous output power cannot exceed 32 W[1]
Accurate?
Program memory
Approx. 2000 program steps, groupable into 99 subroutines[1]
Accurate?
Data memory
Maximum 4095 data values (binary) or 1023 data values (ASCII)[1]
Accurate?
Measurement modes
Plug-in units operate individually or in synchronized modes to perform ten different types of measurements[1]
Accurate?
Self-test and calibration
Self-test at power-up; offset errors in each measurement unit auto-calibrated every 30 minutes[1]
Accurate?
Operating temperature
5 °C to 40 °C[1]
Accurate?
Operating humidity
5% to 80% RH[1]
Accurate?
Storage temperature and humidity
-40 °C to 65 °C; <90% RH at 65 °C[1]
Accurate?
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Integrated measurement functionsCurrent/voltage sources, nanovoltmeters, and picoammeters in one instrument[1]
  • Source Monitor Unit (SMU) sensitivityForces voltage while measuring currents down to 20 fA; forces current while monitoring voltages down to 40 µV[1]
  • HP 41420A High Power SMU2 slots; 40 µV to 200 V, 20 fA to 1 A (±200 V and ±1 A)[1]
  • HP 41421B Medium Power SMU1 slot; 40 µV to 100 V, 20 fA to 100 mA (±100 V and ±100 mA)[1]
  • HP 41422A High Current Unit (HCU)2 slots; pulse mode only; 100 µs pulsed measurements up to ±10 V and 10 A[1]
  • HP 41423A High Voltage Unit (HVU)2 slots; DC or pulsed measurements to ±1000 V; low-leakage design measures currents down to 2 pA[1]
  • HP 41425A Analog Feedback Unit (AFU)1 slot; used with two SMUs for analog voltage/current searches; finds target values in 12 ms[1]
  • Maximum configuration coverage2000 V max using two HVUs; 20 A (pulsed) using two HCUs[1]
  • Maximum common-to-ground voltage±42 V[1]
  • Ground Unit (GNDU)High-current SMU set to 0 V, used as measurement ground; Kelvin connection; maximum offset voltage ±500 µV; current range ±1.6 A[1]
  • Power requirements100/120/220 V ±10%; 240 V -10%/+5%; maximum 750 VA (48-66 Hz)[1]
  • Typical measurement time (SMU)HP 41420A/HP 41421B SMU (20 V/100 mA range, spot measurement): force I or V in 3.5 ms; measure I or V in 4.0 ms[1]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Not publicly documented

The following facts about the 4142 B are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the 4142 B are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • No publicly documented variants, configuration options, or revision breakpoints of the 4142 B are on record.

    Answerable by: an OEM product catalog or an engineer who ordered or specified the tool

  • The control-system platform and OS era of the 4142 B are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the 4142 B are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the 4142 B is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

No research found yet — worked with this tool? Share what you know.